{Ion sensing (EIS) real-time quantitative monitorization of isothermal DNA amplification}

Citation:
Veigas, B., R. Branquinho, J. V. Pinto, P. J. Wojcik, R. Martins, E. Fortunato, and P. V. Baptista, "{Ion sensing (EIS) real-time quantitative monitorization of isothermal DNA amplification}", Biosensors and Bioelectronics, vol. 52: Elsevier, pp. 50–55, 2014.

Abstract:

Field-effect-based devices are becoming a basic structural element in a new generation of microbiosensors. Reliable molecular characterization of DNA and/or RNA is of paramount importance for disease diagnostics and to follow up alterations in gene expression profiles. The use of such devices for point-of-need diagnostics has been hindered by the need of standard or real-time PCR amplification procedures. The present work focuses on the development of a tantalum pentoxide (Ta2O5) based sensor for the real-time label free detection of DNA amplification via loop mediated isothermal amplification (LAMP) allowing for quantitative analysis of the cMYC proto-oncogene. The strategy based on the field effect sensor was tested within a range of 1??108-1011 copies of target DNA, and a linear relationship between the log copy number of the initial template DNA and threshold time was observed allowing for a semi-quantitative analysis of DNA template. The concept offers many of the advantages of isothermal quantitative real-time DNA amplification in a label free approach and may pave the way to point-of-care quantitative molecular analysis focused on ease of use and low cost. ?? 2013 Elsevier B.V.

Notes:

n/a

Related External Link