{Multitechnique description surface analysis system}

Citation:
Silva, J. a M. C., and A. M. C. Moutinho. "{Multitechnique description surface analysis system}." Vaccum. 46 (1995): 1205-1209.

Abstract:

A new multitechnique surface analysis system is presented. It has been designed for research on ion-solid interactions and for survey analysis. SIMS, XPS and AES are the main techniques used. Primary sources are an argon source for standard SIMS and a cesium source for negative and cathionized SIMS, a twin anode (Mg and Al) X-ray source for XPS, and a small spot electron gun for AES and low resolution electron microscopy. The mass spectrometer is a modified quadrupole based probe with an energy analyzer, Dynamic and static SIMS are possible as well as depth profiling. Photoelectrons and Auger electrons are analyzed by a true hemispherical energy analyzer that can also be used for ion spectroscopy. A secondary electron detector is also available. Exchanging the samples is possible through a fast entry air lock. In this small chamber a sputter gun is used to clean the sample. The sample under analysis is supported by a XYZ manipulator and can be temperature controlled in the range 130-850 K